Delivering innovation and workflow efficiencies
Pump tubing-flex System
Made in Germany
The CT motion pump tubing-flex is cleared for 24-hour usage or a maximum of 19 contrast media bottles, whichever comes first.
Patient Tubing
Made in Germany
The patient tubing connect the pump tubing-flex to the cannula on the patient and transports media from the pump tubing-flex to the patient .
Experience the features of Syringeless technology
1. Integrated CM heat retainers to maintain CM temperature in container
2. Media container holder for saline (NaCl). Optional: Saline Rod
3. Media detectors with status indicator
4. Display unit to help support efficient operation
5. Pressure sensors
6. Patient tubing air detector
7. Pump tubing-flex air detector with status indicator
8. Terminal includes injection programs. As well as, safety alerts and timers for day set, contrast media, and patient tubing.
Abbreviations:
CM, Contrast Medium
Tubing System
The tubing system for the ulrichINJECT CT motion consists of pump tubing-flex, spike for CT and patient tubing.
Pump tubing-flex
24-hour multiple patient use
The CT motion pump tubing-flex is cleared for 24-hour usage or a maximum of 19 contrast media bottles, whichever comes first.
1. Swabable valve Connection point for patient tubing to help support aseptic procedure.
2. Check valve
3. Swabable valves Connection point for spikes to help support a hygienic security barrier for the media input (aseptic Global - technique).
4. Pressure sensor unit with integrated particle filter.
Spike for CT (CM/NaCl)
Single-bottle use
Spikes used with Omnipaque™ contrast media in Imaging Bulk Package (IBP) and with the saline (NaCl) container/bag.1. Integrated air filter on spike
Patient tubing
250 cm
Single-patient use
The patient tubing connects the pump tubing-flex to the cannula on the patient and transports media from the pump tubing-flex to the patient.
2. Double check valve systemAbbreviationsIBP, Imaging bulk pack
CM, Contrast medium
NaCl, Saline
Indications for Use
ulrichINJECT CT motion is a contrast media management system that is indicated for the controlled, automatic administration, on the venous side, of contrast media and saline (NaCl), to human subjects undergoing diagnostic examinations in computed tomography (CT) applications.
ulrichINJECT CT motion is specifically indicated for use in CT procedures for the delivery of Omnipaque™ (Iohexol) Injection, solution - GE Healthcare Inc. contrast media as supplied in Imaging Bulk Packages (IBP).
Pump tubing-flex is used for a maximum time of twenty four (24) hours. When used with Omnipaque™ IBP, a maximum of 19 bottles of contrast media can be used or maximum time of twenty four (24) hours of Pump tubing-flex, or whichever comes first, with a maximum of eight (8) hours per contrast media or saline container.
Spike for CT disposable is for single-bottle use only and must be discarded with the media container. The Patient tubing must be discarded after each patient procedure.
ulrichINJECT CT motion is to be used only by and under quasi-continuous supervision of trained healthcare professionals in an appropriate licensed healthcare facility, in a room designated for radiological procedures that involve intravascular administration of contrast agent.